Product Info Summary
| SKU: | EK1834 |
|---|---|
| Size: | 96 wells/kit, with removable strips. |
| Reactive Species: | Human |
| Application: | ELISA |
| Sample Types: | cell culture supernatants, serum and plasma (heparin). |
Product info
Product Name
Human REG3A ELISA Kit PicoKine®
SKU/Catalog Number
EK1834
Size
96 wells/kit, with removable strips.
*Question: How many samples can I assay/run in this kit?
Description
Human REG3A ELISA Kit PicoKine® (96 Tests). Quantitate Human REG3A in cell culture supernatants, serum and plasma (heparin). Sensitivity: 0.1ng/ml. The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.
Storage & Handling
Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
Cite This Product
Human REG3A ELISA Kit PicoKine® (Boster Biological Technology, Pleasanton CA, USA, Catalog # EK1834)
Clonality of Antibodies
See Datasheet for details
Standard Protein
Expression system for standard: NS0; Immunogen sequence: E27-D175
Capture/Detection Antibodies
The capture antibody is monoclonal antibody from mouse and the detection antibody is polyclonal antibody from goat.
Sensitivity
<0.1 ng/ml
Assay Range
0.78 ng/ml - 50 ng/ml
Standard Dilution Instructions

See datasheet of EK1834 for more details
Cross-reactivity
There is no detectable cross-reactivity with other relevant proteins.
Reactive Species
EK1834 is reactive to REG3A in Human samples
Validated Sample Types
cell culture supernatants, serum and plasma (heparin).
Application Guarantee
EK1834 is guaranteed for ELISA in Human by Boster Guarantee
See how Boster Bio validate our ELISA kits: ELISA Validation Information
Background of REG3A
Regenerating islet-derived protein 3 alpha (or Regenerating islet-derived protein III-alpha) formerly known as HIP/PAP (Hepatocarcinoma-Intestine-Pancreas/Pancreatitis-Associated Protein) and peptide 23 is a protein that in humans is encoded by the REG3A gene. This gene is mapped to 2p12. It encodes a pancreatic secretory protein that may be involved in cell proliferation or differentiation. It has similarity to the C-type lectin superfamily. The enhanced expression of this gene is observed during pancreatic inflammation and liver carcinogenesis. The mature protein also functions as an antimicrobial protein with antibacterial activity. Alternate splicing results in multiple transcript variants that encode the same protein.
Kit Components
| Catalog Number | Description | Quantity |
|---|---|---|
| EK1834-CAP | Anti-Human REG3A Pre-coated 96-well strip microplate | 1 |
| EK1834-ST | Human REG3A Standard | 2 vials, 50 ng/tube |
| EK1834-DA | Human REG3A Biotinylated antibody (100x) | 100ul |
| AR1103 | Avidin-Biotin-Peroxidase Complex (100x) | 100ul |
| AR1106-1 | Sample Diluent | 30ml |
| AR1106-2 | Antibody Diluent | 12ml |
| AR1106-3 | Avidin-Biotin-Peroxidase Diluent | 12ml |
| AR1104 | Color Developing Reagent (TMB) | 10ml |
| AR1105 | Stop Solution | 10ml |
| AR1106-5 | Wash Buffer (25x) | 20ml |
| PLA-SEA | Adhesive plate sealers | 4 |
*The kit components are not available for individual purchase.
Materials Required But Not Included In Kit
- Microplate Reader capable of reading absorbance at 450nm.
- Incubator.
- Automated plate washer (optional).
- Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions.
- Multichannel pipettes are recommended for large amount of samples.
- Deionized or distilled water.
- 500ml graduated cylinders.
- Test tubes for dilution.
Data Examples, Quality Control Data & Sample Dilution
Validation Standard Curve O.D. At 450nm
| Concentration (ng/ml) | 0 | 0.78 | 1.56 | 3.12 | 6.25 | 12.5 | 25 | 50 |
| O.D. | 0.078 | 0.148 | 0.189 | 0.261 | 0.456 | 0.762 | 1.223 | 1.982 |
Data Example Images
Click image to see more details
Human REG3A PicoKine ELISA Kit standard curve
Recommended Sample Dilution Ratios
According to our internal validation assays using this ELISA kit, to detect REG3A, Dilution ratio of 1:50, concentration in serum and plasma is 300-500 ng/ml..
Intra Assay Consistency & Inter Assay Consistency
We measured random samples of Human REG3A ELISA Kit PicoKine® within the same batch/lot to ensure the consistency of the kits' performances. ELISA intra assay consistency is measured using wells from the same plate/assay kit. ELISA inter assay consistency is measured using wells from different plates from the same batch production/lot.
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 16 | 16 | 16 | 24 | 24 | 24 |
| Mean (pg/ml) | 1176 | 6316 | 19450 | 1433 | 6401 | 19872 |
| Standard deviation | 58.80 | 385.28 | 836.35 | 65.92 | 409.66 | 1430.78 |
| CV (%) | 5.0% | 6.1% | 4.3% | 4.6% | 6.4% | 7.2% |
Specific Publications For Human REG3A ELISA Kit PicoKine® (EK1834)
Loading publications
Recommended Resources
Here are featured tools and databases that you might find useful.
- Boster's Pathways Library
- Protein Databases
- Bioscience Research Protocol Resources
- Data Processing & Analysis Software
- Photo Editing Software
- Scientific Literature Resources
- Research Paper Management Tools
- Molecular Biology Software
- Primer Design Tools
- Bioinformatics Tools
- Phylogenetic Tree Analysis
Customer Reviews
0 Reviews For Human REG3A ELISA Kit PicoKine®
Customer Q&As
Have a question?
Find answers in Q&As, reviews.
Can't find your answer?
Submit your question
5 Customer Q&As for Human REG3A ELISA Kit PicoKine®
Question
Q: if the enzyme conjugated REG3A antibodies are mixed with the substrate, will that convert the substrate into the enzymatic reaction product? Or the enzyme function is only activated when the antibody is attached with the REG3A antigen?
Verified Customer
Verified customer
Asked: 2018-11-15
Answer
A: The enzyme is always active. Avoid contaminating the substrate with enzyme prior to the incubation otherwise it compromises the assay with false positive signal.
Boster Scientific Support
Answered: 2018-11-15
Question
Q: which procedure should I follow in order to thaw whole blood sample for REG3A ELISA after freezing?
Verified Customer
Verified customer
Asked: 2018-04-23
Answer
A: do not freeze and thaw whole blood. erythrocytes are fragile and, if frozen and thawed, will undergo hemolysis rendering the samples useless. To keep your blood samples to test REG3A for a later time, you should let the blood clot in glass tubes and separate the serum to freeze for later analysis.
Boster Scientific Support
Answered: 2018-04-23
Question
Q: Can REG3A ELISA Kits be used with tissue homogenates (or other non-validated sample types)?
Verified Customer
Verified customer
Asked: 2017-06-29
Answer
A: Unfortunately, Boster Bio has not routinely validated tissue homogenates as a sample type for ELISA kits. This does not mean that ELISA kits are not suitable for other sample types than we have tested: it means further investigation is a must. One will need to perform a spike and recovery study to determine if an unvalidated sample type will work with a particular kit. To perform a spike and recovery experiment, one should divide a sample into two aliquots. In one of the aliquots, the user should spike in a known amount of the kit standard. a dilution series is performed comparing the spiked versus the unspiked sample. Generally, samples with expected recovery and linearity between 80-120% are considered acceptable. This method can be used to validate any sample type that has not been assessd by Boster Bio. for a more detailed spike and recovery protocol, please contact technical support.
Note: acceptable ranges should be determined individually by each laboratory. Additionally, technical support can help determine if a buffer component is not compatible with a given ELISA kit. please see the Citations tab on the product webpage for peer-reviewed papers utilizing a wide range of sample types. We also have an innovator's reward program where if the user validates our ELISA kits in applications or samples previously not validated by Boster Bio or other users, and share such information with us by submit a review, we will reward the user's efforts with a free antibody or ELISA kit from our catalog. Biocompare.com will also give $20 Amazon giftcard as an additional reward, if the review is submitted there as well.
Boster Scientific Support
Answered: 2017-06-29
Question
Q: Are Boster Bio recombinant proteins and antibodies sterile?
Verified Customer
Verified customer
Asked: 2017-04-29
Answer
A: although the vials are bottled using aseptic techniques, heat-treated vials, and sterile stock solutions, they are not considered or guaranteed to be sterile. If sterile material is a must for an experiment, the material can be filtered through a 0.2 micron filter designed for use with biological fluids.
Boster Scientific Support
Answered: 2017-04-29
Question
Q: What is the optimal O.D. value for REG3A ELISA kit? I used your REG3A ELISA on serum samples. For my positive control, I received an O.D. value of 0.826, while my negative control received a value of 0.136. I obtained both of these controls from the ELISA kit, where your kit's typical data shows O.D. values much higher than my positive control and your background is lower. My samples O.D. values are around 0.225 and the highest is only 0.357. is it safe to say these samples contain REG3A even though the O.D. values are not very high?
Verified Customer
Verified customer
Asked: 2016-10-27
Answer
A: The absolute O.D. values may change according to incubation time. The more you incubate the higher the O.D. values are going to be. a point of focus should be is whether your sample O.D. values are statistically significantly higher than your blank values. considering your assay, you could extend your development time in the substrate incubation step to obtain higher O.D. values, as long as your negative controls' O.D. values are not increasing faster in proportion to your positive controls. normally, a sample with O.D. value 2 standard deviations higher than your negative controls can be considered positive. We calculate the sensitivity of this ELISA kit by changeing cutoff O.D. value, calculated as the average of 20 negative controls plus 2 standard deviations of the 20 negative controls, into a concentration. in other words, when we claim this REG3A ELISA kit to have sensitivity of 0.1ng/ml, that means the minimum amount of REG3A that can be declared/interpreted as positive by the above standard is 0.1ng/ml.
Boster Scientific Support
Answered: 2016-10-27