Product Info Summary
| SKU: | EK0412 |
|---|---|
| Size: | 96 wells/kit, with removable strips. |
| Reactive Species: | Rat |
| Application: | ELISA |
| Sample Types: | cell culture supernatants, serum and plasma (heparin, EDTA, citrate). |
Product info
Product Name
Rat IL-6/Interleukin-6 ELISA Kit PicoKine®
SKU/Catalog Number
EK0412
Size
96 wells/kit, with removable strips.
*Question: How many samples can I assay/run in this kit?
Description
Rat IL-6/Interleukin-6 ELISA Kit PicoKine® (96 Tests). Quantitate Rat Il6 in cell culture supernatants, serum and plasma (heparin, EDTA, citrate). Sensitivity: 5pg/ml. The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.
Storage & Handling
Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
Cite This Product
Rat IL-6/Interleukin-6 ELISA Kit PicoKine® (Boster Biological Technology, Pleasanton CA, USA, Catalog # EK0412)
Clonality of Antibodies
See Datasheet for details
Standard Protein
Expression system for standard: E.coli; Immunogen sequence: F25-T211
Sensitivity
<5 pg/ml
Assay Range
62.5 pg/ml - 4,000 pg/ml
Standard Dilution Instructions

See datasheet of EK0412 for more details
Cross-reactivity
There is no detectable cross-reactivity with other relevant proteins.
Reactive Species
EK0412 is reactive to Il6 in Rat samples
Validated Sample Types
cell culture supernatants, serum and plasma (heparin, EDTA, citrate).
Application Guarantee
EK0412 is guaranteed for ELISA in Rat by Boster Guarantee
See how Boster Bio validate our ELISA kits: ELISA Validation Information
Background of Il6
Interleukin-6 (IL-6) is a protein that in humans is encoded by the IL6 gene. Interleukin-6 (IL-6) is involved not only in the hepatic acute phase response but also in adipose tissue metabolism, lipoprotein lipase activity, and hepatic triglyceride secretion. IL-6 is a pleiotropic cytokine that plays a critical role in bone resorption. IL-6 has been suggested to stimulate the HPA axis during immune activation independent of the input of hypothalamic corticotropin-releasing hormone (CRH). Interleukin-6 (IL-6) is an immunoregulatory cytokine that activates a cell-surface signaling assembly composed of IL-6, the IL-6 alpha-receptor (IL-6Ralpha), and the shared signaling receptor gp130. The standard product used in this kit is recombinant rat IL-6, consisting of 188 amino acids with the molecular mass of 21.8kDa.
Kit Components
| Catalog Number | Description | Quantity |
|---|---|---|
| EK0412-CAP | Anti-Rat Il6 Pre-coated 96-well strip microplate | 1 |
| EK0412-ST | Rat Il6 Standard | 2 vials, 10 ng/tube |
| EK0412-DA | Rat Il6 Biotinylated antibody (100x) | 100ul |
| AR1103 | Avidin-Biotin-Peroxidase Complex (100x) | 100ul |
| AR1106-1 | Sample Diluent | 30ml |
| AR1106-2 | Antibody Diluent | 12ml |
| AR1106-3 | Avidin-Biotin-Peroxidase Diluent | 12ml |
| AR1104 | Color Developing Reagent (TMB) | 10ml |
| AR1105 | Stop Solution | 10ml |
| AR1106-5 | Wash Buffer (25x) | 20ml |
| PLA-SEA | Adhesive plate sealers | 4 |
*The kit components are not available for individual purchase.
Materials Required But Not Included In Kit
- Microplate Reader capable of reading absorbance at 450nm.
- Incubator.
- Automated plate washer (optional).
- Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions.
- Multichannel pipettes are recommended for large amount of samples.
- Deionized or distilled water.
- 500ml graduated cylinders.
- Test tubes for dilution.
Data Examples, Quality Control Data & Sample Dilution
Validation Standard Curve O.D. At 450nm
| Concentration (pg/ml) | 0 | 62.5 | 125 | 250 | 500 | 1000 | 2000 | 4000 |
| O.D. | 0.051 | 0.117 | 0.194 | 0.296 | 0.525 | 0.919 | 1.523 | 1.982 |
Data Example Images
Click image to see more details
Rat IL-6 PicoKine ELISA Kit standard curve
Recommended Sample Dilution Ratios
According to our internal validation assays using this ELISA kit, to detect Il6, Dilution ratio of 1:1, concentration in serum and plasma is less than the lowest standard, 62.5 pg/ml..
Some articles we found to cite concentrations of Il6 in samples: 23326132 (Pubmed IDs).
Intra Assay Consistency & Inter Assay Consistency
We measured random samples of Rat IL-6/Interleukin-6 ELISA Kit PicoKine® within the same batch/lot to ensure the consistency of the kits' performances. ELISA intra assay consistency is measured using wells from the same plate/assay kit. ELISA inter assay consistency is measured using wells from different plates from the same batch production/lot.
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 16 | 16 | 16 | 24 | 24 | 24 |
| Mean (pg/ml) | 193 | 579 | 1531 | 210 | 564 | 1466 |
| Standard deviation | 9.65 | 35.89 | 90.32 | 11.55 | 40.04 | 109.95 |
| CV (%) | 5% | 6.2% | 5.9% | 5.5% | 7.1% | 7.5% |
Specific Publications For Rat IL-6/Interleukin-6 ELISA Kit PicoKine® (EK0412)
Loading publications
Recommended Resources
Here are featured tools and databases that you might find useful.
- Boster's Pathways Library
- Protein Databases
- Bioscience Research Protocol Resources
- Data Processing & Analysis Software
- Photo Editing Software
- Scientific Literature Resources
- Research Paper Management Tools
- Molecular Biology Software
- Primer Design Tools
- Bioinformatics Tools
- Phylogenetic Tree Analysis
Customer Reviews
0 Reviews For Rat IL-6/Interleukin-6 ELISA Kit PicoKine®
Customer Q&As
Have a question?
Find answers in Q&As, reviews.
Can't find your answer?
Submit your question
10 Customer Q&As for Rat IL-6/Interleukin-6 ELISA Kit PicoKine®
Question
Q: which procedure should I follow in order to thaw whole blood sample for IL-6 ELISA after freezing?
Verified Customer
Verified customer
Asked: 2020-05-10
Answer
A: we do not recommend freezing and thaw whole blood. erythrocytes are fragile and, if frozen and thawed, will undergo hemolysis rendering the samples useless. To keep your blood samples to test IL-6 for a later time, you should let the blood clot in glass tubes and collect the serum to freeze for later use.
Boster Scientific Support
Answered: 2020-05-10
Question
I would like to know more about Rat IL-6 ELISA Kit antibody. Could this be used also as neutralizing antibody? As it stated in the product notes that this antibody can be used to block IL-6/IL-6R interactions. You think it is just the same as neutralizing antibody?
Verified Customer
Verified customer
Asked: 2020-03-19
Answer
The Rat IL-6 ELISA Kit antibody can be used to prevent IL-6 binding to its receptor, therefore it could be used in a neutralizing type of assay.
Boster Scientific Support
Answered: 2020-03-19
Question
To Technical Support: I would like to inquire if this Rat IL-6 ELISA Kit can be used for urine samples?
Verified Customer
Verified customer
Asked: 2020-01-08
Answer
The Rat IL-6 ELISA Kit has been successfully tested with urine samples, we can guarantee that it will worked well. However, we would recommend to start a minimal dilution (such as 2-fold) since based on our experience cytokine levels are often very low in urine. Here
Boster Scientific Support
Answered: 2020-01-08
Question
Q: how to analyze ELISA data? I measured IL-6 level in plasma.
W. Carter
Verified customer
Asked: 2019-11-09
Answer
A: please read this article on ELISA data analysis. bosterbio.com/elisa-data-analysis-instructions. we also provide a convenient online tool that you can use to analyze ELISA data. bosterbio.com/biology-research-tools/elisa-data-analysis-online
Boster Scientific Support
Answered: 2019-11-09
Question
Is the Rat IL-6 PicoKine
Verified Customer
Verified customer
Asked: 2019-06-04
Answer
We haven't tested
Boster Scientific Support
Answered: 2019-06-04
Question
Q: can I use citrate plasma as samples in Rat IL-6 Picokine® ELISA Kit (Catalog # EK0412)?
Verified Customer
Verified customer
Asked: 2019-03-18
Answer
A: Chelating agents such as EDTA, Heparin and Citrate can irreversibly bind metal ions from the functional domain of IL-6 causing degradation of its protein structure. IL-6 may be denatured as a result and may compromise the assay's measurements. The chilating sites could also be too close to the epitopes required for detection and block the antigen antibody reaction. We have tested the IL-6 ELISA, treating samples with various anticoagulants and decided that heparin, EDTA or citrate can be used for treatment of blood/plasma samples. Do not use other anticoagulents when collecting samples.
Boster Scientific Support
Answered: 2019-03-18
Question
Q: how many samples can be assayed in a Picokine® ELISA Kit?
Verified Customer
Verified customer
Asked: 2019-02-25
Answer
A: The Picokine® ELISA Kits will generally run a 7-point standard curve, non-specific binding wells, and 39 samples in duplicate. this may might differ by kit so please refer to each datasheet for details.
Boster Scientific Support
Answered: 2019-02-25
Question
Can EK0412 be used with cytosolic fractions?
Verified customer
Asked: 2019-02-08
Answer
We haven't tested tissue lysates/cytosolic fractions with Rat IL-6 ELISA Kit PicoKine EK0412. Endogenous biotin in tissue homogenates might introduce false positive signal. We recommend to run a pilot test to confirm the effect of endogeous biotin first. Add tissue homogenates into the wells then add ABC and TMB without adding any biotinylated detection antibody to see if any signal will be observed. If no signal is produced, proceed and work on the tissue sample by using the kit.
Boster Scientific Support
Answered: 2019-02-08
Question
Q: how are cell lysates prepared for use in Picokine® ELISA kits?
Verified Customer
Verified customer
Asked: 2018-11-16
Answer
A: in those Picokine® ELISAs where cell or tissue lysate is a validated sample type, sample preparation instructions for lysate are included in the product insert. Components in lysate and lysis buffer can have an affect on immunoreactivity, so if lysate is not a validated sample type, care must be taken in sample preparation and validation.
Boster Scientific Support
Answered: 2018-11-16
Question
Q: What is the optimal O.D. value for IL-6 ELISA kit? I performed your IL-6 ELISA on serum samples. For my positive control, I received an O.D. value of 0.826, while my negative control received a value of 0.136. I obtained both of these controls from the ELISA kit, where your kit's typical data shows O.D. values much higher than my positive control and your background is lower. My samples O.D. values are around 0.225 and the highest is only 0.357. can I consider these samples contain IL-6 even though the O.D. values are not very high?
Verified Customer
Verified customer
Asked: 2015-04-07
Answer
A: The absolute O.D. values may change according to incubation time. The more you incubate the higher the O.D. values are going to be. a point of focus should be is whether your sample O.D. values are statistically significantly higher than your blank values. regarding your protocol, you could extend your development time in the substrate incubation step to obtain higher O.D. values, as long as your negative controls' O.D. values are not increasing faster in relation to your positive controls. normally, a sample with O.D. value 2 standard deviations higher than your negative controls can be considered positive. We calculate the sensitivity of this ELISA kit by converting cutoff O.D. value, calculated as the average of 20 negative controls plus 2 standard deviations of the 20 negative controls, into a concentration. in other words, when we claim this IL-6 ELISA kit to have sensitivity of 5pg/ml, that means the minimum amount of IL-6 that can be declared/interpreted as positive by the above standard is 5pg/ml.
Boster Scientific Support
Answered: 2015-04-07