Product Info Summary
| SKU: | EK1458 |
|---|---|
| Size: | 96 wells/kit, with removable strips. |
| Reactive Species: | Mouse |
| Application: | ELISA |
| Sample Types: | cell culture supernatants, serum, plasma (heparin, EDTA) and cell lysates. |
Product info
Product Name
Mouse FABP4/A Fabp ELISA Kit PicoKine®
SKU/Catalog Number
EK1458
Size
96 wells/kit, with removable strips.
*Question: How many samples can I assay/run in this kit?
Description
Mouse FABP4/A Fabp ELISA Kit PicoKine® (96 Tests). Quantitate Mouse Fabp4 in cell culture supernatants, serum, plasma (heparin, EDTA) and cell lysates. Sensitivity: 15pg/ml. The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.
Storage & Handling
Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
Cite This Product
Mouse FABP4/A Fabp ELISA Kit PicoKine® (Boster Biological Technology, Pleasanton CA, USA, Catalog # EK1458)
Clonality of Antibodies
See Datasheet for details
Standard Protein
Expression system for standard: E.coli; Immunogen sequence: M1-A132
Sensitivity
<15 pg/ml
Assay Range
156 pg/ml - 10,000 pg/ml
Standard Dilution Instructions

See datasheet of EK1458 for more details
Cross-reactivity
There is no detectable cross-reactivity with other relevant proteins.
Reactive Species
EK1458 is reactive to Fabp4 in Mouse samples
Validated Sample Types
cell culture supernatants, serum, plasma (heparin, EDTA) and cell lysates.
Application Guarantee
EK1458 is guaranteed for ELISA in Mouse by Boster Guarantee
See how Boster Bio validate our ELISA kits: ELISA Validation Information
Background of Fabp4
FABP4, also known as aP2 (adipocyte Protein 2), encodes the fatty acid binding protein found in adipocytes. The FABP4 gene is mapped to 8q21. FABP4 functions as a cytosolic lipid chaperone in macrophages and is involved in regulating macrophage endoplasmic reticulum (ER) stress. Aditionally, FABP4 protein is expressed in HBEs and is strongly upregulated by both IL4 and IL13 and downregulated by IFN-gamma, suggesting a role in allergic airway inflammation.
Kit Components
| Catalog Number | Description | Quantity |
|---|---|---|
| EK1458-CAP | Anti-Mouse Fabp4 Pre-coated 96-well strip microplate | 1 |
| EK1458-ST | Mouse Fabp4 Standard | 2 vials, 10 ng/tube |
| EK1458-DA | Mouse Fabp4 Biotinylated antibody (100x) | 100ul |
| AR1103 | Avidin-Biotin-Peroxidase Complex (100x) | 100ul |
| AR1106-1 | Sample Diluent | 30ml |
| AR1106-2 | Antibody Diluent | 12ml |
| AR1106-3 | Avidin-Biotin-Peroxidase Diluent | 12ml |
| AR1104 | Color Developing Reagent (TMB) | 10ml |
| AR1105 | Stop Solution | 10ml |
| AR1106-5 | Wash Buffer (25x) | 20ml |
| PLA-SEA | Adhesive plate sealers | 4 |
*The kit components are not available for individual purchase.
Materials Required But Not Included In Kit
- Microplate Reader capable of reading absorbance at 450nm.
- Incubator.
- Automated plate washer (optional).
- Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions.
- Multichannel pipettes are recommended for large amount of samples.
- Deionized or distilled water.
- 500ml graduated cylinders.
- Test tubes for dilution.
Data Examples, Quality Control Data
Validation Standard Curve O.D. At 450nm
| Concentration (pg/ml) | 0 | 156 | 312 | 625 | 1250 | 2500 | 5000 | 10000 |
| O.D. | 0.019 | 0.051 | 0.087 | 0.156 | 0.296 | 0.611 | 1.185 | 1.976 |
Data Example Images
Click image to see more details
Mouse FABP4 PicoKine ELISA Kit standard curve
Intra Assay Consistency & Inter Assay Consistency
We measured random samples of Mouse FABP4/A Fabp ELISA Kit PicoKine® within the same batch/lot to ensure the consistency of the kits' performances. ELISA intra assay consistency is measured using wells from the same plate/assay kit. ELISA inter assay consistency is measured using wells from different plates from the same batch production/lot.
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 16 | 16 | 16 | 24 | 24 | 24 |
| Mean (pg/ml) | 344 | 1350 | 4564 | 359 | 1418 | 4994 |
| Standard deviation | 26.83 | 74.25 | 351.42 | 29.07 | 93.58 | 409.5 |
| CV (%) | 7.8% | 5.5% | 7.7% | 8.1% | 6.6% | 8.2% |
Specific Publications For Mouse FABP4/A Fabp ELISA Kit PicoKine® (EK1458)
Loading publications
Recommended Resources
Here are featured tools and databases that you might find useful.
- Boster's Pathways Library
- Protein Databases
- Bioscience Research Protocol Resources
- Data Processing & Analysis Software
- Photo Editing Software
- Scientific Literature Resources
- Research Paper Management Tools
- Molecular Biology Software
- Primer Design Tools
- Bioinformatics Tools
- Phylogenetic Tree Analysis
Customer Reviews
1 Reviews For Mouse FABP4/A Fabp ELISA Kit PicoKine®
No Complaints. A Simple, Effective ELISA from BosterBio--David Nelson, University of Wisconsin, Research Scientist
Excellent

Source: Biocompare.com
| SKU | EK1458 |
|---|---|
| Application | ELISA |
| Sample | Mouse plasma (with EDTA as anticoagulant) |
| Detection | -- |
"Reagents are easy to prep or already 1X. Two vials of standard allows for breaking the kit into two assays. 10 ml of TMB and Stop solution is cutting it close, but was not an issue. Really nothing was bad. Standard curve was easy to make, and 4-parameter curve fit perfect (R2>0.99). CV% for replicates on standard curve were a bit higher than I expected - around 5-10% - but consistent with the product literature. I'd recommend the kit. I did the assay at room temperature (instead of the recommended 37°C) because I didn't want to deal with going back and forth into the room that houses our 37°C incubator. Doing the assay at room temperature takes a bit longer (for the incubations). The data looked good (based on R2 for standard curve and CV% for replicates. In the image, Columns A and B are the standard curve." (Customer’s Tip: "100-fold dilution of mouse plasma resulted in all data falling on standard curve.")
David Nelson
Verified customer
Submitted 2019-12-06
Submit your question